The Actiwatch Spectrum actigraphy system incorporate high-quality biocompatible materials and well-engineered sensors - accelerometer, light and off wrist to provide 24/7wearability and accurate data collection. The Actiwatch Spectrum Starter kit (actiwatch Spectrum, communication dock and actiwave software) includes everything to get started with actigraphy.
Features
- Comfortable, aesthetic design
- Off-wrist detector
- Time and date indicators (like an actual wristwatch)
- Device status indicator (logging, not logging, on wrist, off wrist, successful marker push)
- Seven Data Channels: white photopic light illuminance, red, green and blue light irradiance, activity, off-wrist and event marking
- Moisture protection and increased resistance to environmental challenges (moisture, shock, dust, etc.)
- Backlit display Infrared communication to the dock/reader
- One-year battery life – changeable at factory
- Large memory (2 MBit Non-volatile)
- Faster communications and data-download rate
Specifications
| |
Size |
|
48mm x 37mm x 14mm |
|
| |
Weight |
|
30 grams (with band) |
|
| |
Case |
|
ABS Blend (Biocompatible) |
|
| |
Wrist Band |
|
BASF Elastollan 1175A10W (TPU) with titanium buckle |
|
| |
Hospital Bands |
|
Accomodates 7mm hospital band |
|
| |
Battery type |
|
CR 2430 Lithium Coin Cell (factory replaced) |
|
| |
Battery Charge Time |
|
6 hours empty to full |
|
| |
Memory |
|
Non volitile 2 Mbits |
|
| |
Waterproof |
|
1m for 30 min per IPX7 IEC 60529 |
|
| |
Operating Temperature Range |
|
5oC to 40oC |
|
| |
Resistance |
|
Resists dust, water, heat, perspiration |
|
| |
|
|
|
|
| |
|
|
Epoch Length |
|
| |
|
|
15 sec |
30 sec |
1 min |
|
| |
Epochs per day |
|
5760 |
2280 |
1440 |
|
| |
Logging time (activity only) |
|
45.5 days |
91 days |
182 days |
|
| |
Logging time (activity and RGB light) |
|
9 days* |
18 day* |
36 days * |
|
| |
|
|
* also records 2 and 5 min epochs with white light and RGB channels
|
|
| |
|
|
|
|
| |
Off Wrist Dectector |
|
|
|
| |
Sensor |
|
Capacitive |
|
| |
On-Wrist Activation Range |
|
< 2mm |
|
| |
|
|
|
|
| |
Accelerometer |
|
|
|
| |
Type |
|
Solid state "Piezo-electric" accelerometer |
|
| |
Range |
|
0.5 - 2 G peak value |
|
| |
Bandwidth |
|
0.35 - 7.5 Hz Typical |
|
| |
Sensitivity |
|
0.025 G (at 2 count level) |
|
| |
Sampling Rate |
|
32 Hz |
|
| |
|
|
|
|
| |
Light Sensor |
|
|
|
| |
Type |
|
Colour sensitive Photodiode |
|
| |
Wavelength Range |
|
400 -900 nanometers |
|
| |
Measurement Units |
|
Irradiance (microwatts/cm2)
Photon Flux (photons/cm2/sec)
Photopic illuminance in Lumens/m2 (Lux) |
|
| |
Illuminance Range |
|
5 -100,000 Lux |
|
| |
Accuracy |
|
10 % at 3000 Lux (typical) |
|